NOTE
External equipment intended for connection to signal input, signal output or
other connectors, shall comply with relevant IEC standard (e.g. IEC 60950
for IT equipment and the IEC 60601 series for medical electrical equipment).
In addition, all such combinations - systems - shall comply with the IEC
60601-1 standard. Equipment not complying to IEC 60601-1 shall be kept
outside the patient area.
Any person who connects external equipment to signal input, signal output
or other connectors has formed a system and is therefore responsible for the
system to comply with the requirements of IEC 60601-1. If in doubt, contact
a qualified technician or your local representative.
7.2 Safety switches
Care should be taken when driving the patient chair from one position to
another. Obstructions in the patient chair’s line of movement activate safety
switches that stop the motorised movements. The safety switches and their
functions are described below.
1. Backrest
An obstruction between the backrest and the floor when driving down the
chair and/or the backrest stops downwards chair and backrest
movements. Remove the obstruction to resume normal operation.
2. Chair
An obstruction between the chair and the floor stops downwards chair
and backrest movements. Remove the obstruction to resume normal
operation.
3. Legrest
An obstruction between the legrest and the floor stops downwards chair
and backrest movements. Remove the obstruction to resume normal
operation. The location of the safety switch depends on which legrest
your dental unit is equipped with.
4. OP delivery arm joint
An obstruction between the OP delivery arm and the chair when driving
up the chair stops upwards chair movements. Remove the obstruction to
resume normal operation.
5. OP delivery arm
An obstruction between the OP delivery arm and the chair when driving
up/down the chair stops up-/downwards chair movements. Remove the
obstruction to resume normal operation.
7 For your safety
User's manual
Planmeca Sovereign Classic 15
Summary of Contents for Sovereign Classic
Page 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Page 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Page 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Page 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Page 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Page 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Page 296: ......
Page 297: ......