Planmeca ProSensor 21
TECHNICAL SPECIFICATIONS
User’s Manual
A.2
Preprogrammed settings values
For programming these values see the Planmeca ProX
User’s manual, section 16.3 “Programming the
preprogrammed settings”.
NOTE Two sets of exposure values (time/kV/mA) have been
programmed for each exposure region: one for adult
mode and one for child mode.
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV
mA
time
kV
mA
time
kV
mA
time
Adult
Maxilla
60
7
0.100
63
6
0.1
63
6
0.125
Mandible 60
7
0.08
63
6
0.08
63
6
0.100
Child
Maxilla
60
7
0.063
60
7
0.08
60
7
0.100
Mandible 60
7
0.05
60
7
0.064
60
7
0.080
OCCLUSAL
EXPOSURE
ENDODONTIC
BITE-WING
kV
mA
time
kV
mA
time
kV
mA
time
Adult
Maxilla
70
6
0.100
60
7
0.100
60
7
0.100
Mandible 70
6
0.100
Child
Maxilla
66
6
0.080
60
7
0.080
60
7
0.080
Mandible 66
6
0.080