Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-71
CEPHALOSTAT ADJUSTMENT
Technical Manual
7.5
Checking the head support position
Place the beam alignment tool to the sensor alignment tool to horizontal position (see figure
below).
Figure 134
Enter the ceph exposure mode. The rotating unit will move automatically to the correct posi-
tion for taking cephalometric exposures. Select collimator number 6.
Figure 135
Beam alignment tool