Planmeca Proline EC X-ray unit C-1
Chapter
Technical Manual
C
HELP & ERROR MESSAGES
1 HELP MESSAGES
The following is a list of the help messages. These messages appear on the main display if a
key is pressed that has no function, or if the function is not allowed for some reason. The help
message disappear automatically when you release the key.
HELP CODE
HELP MESSAGE EXPLANATION
HE 0
The key has no function in this mode.
HE 1
The key has no function with the control key.
HE 2
The error code must first be cleared from the display by pressing the
control key.
HE 6
The temple support motor has been inactivated. The temple support
movement does not operate.
HE 8
The child mode can not be selected in double TMJ exposure mode.
HE 10
The X-ray unit is busy, wait until the previous function is completed.
HE 15
The X-ray unit is not communicating with the PC, e.g. the Dimaxis soft-
ware is not running.
HE 16
The patient is not selected in the Dimaxis software.
HE 17
The sensor head is not in its position.
HE 18
In the cephalometric mode there is also a sensor head in the rotating
unit in addition to the sensor head attached to the cephalostat. Remove
the sensor head from the rotating unit.
HE 19
The unit exposure mode (panoramic/cephalometric) differs from the
mode selected in the Dimaxis software.
Summary of Contents for Proline EC ceph
Page 2: ......
Page 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Page 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Page 208: ......
Page 209: ......