7.6 Replacing water or air main valve assembly
CAUTION
Ensure that the water and air supply lines are turned off.
a)
Decompress the water and air pressures as described in section 7.1 “Decompressing water
and air pressures” on page G-36.
b)
Disconnect the control cable from the main valve you are working on.
c)
Detach the locking nut from the main valve. Remove the old main valve. Detach the valve
bottom plate.
d)
Remove the valve support and its attachment plate by unscrewing the four screws. Remove
the O-ring, spring and core.
e)
Install the new main valve assembly in reverse order.
f)
Check that the valve functions correctly.
Bmain
v
alv
es1202.eps
Air main valve
Water main valve
Valve cable
(blue leads)
Valve cable
(green leads)
Locking nut
Core
Valve support
Screws
Air main valve assembly
Valve bottom plate
Support attachment plate
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......