Chapter G - PARTS REPLACEMENT & REPAIR
Planmeca Compact i G-21
ELECTRICAL PARTS REPLACEMENT
Technical Manual
5.2 Replacing fuses
WARNING
Always replace fuses with the same type and rating.
Ratings for Mains Fuse Neutral (F11) and Mains Fuse Live (F12)
NOTE
The mains fuse ratings depend on the selected voltage setting.
Fuse Panel
Code
Fuse Description
Fuse Rating & Type
F5
Operating light fuse
T6.3AL, slow blow
F6
Separator 24V~ fuse
T8AL, slow blow
F7
Water heater 24V~ fuse
T6.3AL, slow blow
F8
Syringe heater 24V~ fuse
T6.3AL, slow blow
F9
Elect24V Fuse A
T4AL, slow blow
F10
Elect24V Fuse B
T4AL, slow blow
Mains
Voltage
Setting
Fuse Rating & Type
100V
~
Schurter T12.5A, slow blow,
0034.3128
115V
~
Schurter T12.5A, slow blow,
0034.3128
220-240V
~
T6.3A, slow blow