Chapter G - PARTS REPLACEMENT & REPAIR
G-14 Planmeca Compact i
SUCTION SYSTEM RELATED
Technical Manual
4 SUCTION SYSTEM RELATED
4.1 Function of the PM Microvac II separating tank
Idle situation
Suction hose removed from the holder
Electrical function:
No control voltage is connected to any of the valves, suc-
tion motor stopped (if controlled by the unit).
Pneumatical function:
All the membranes are open.
Electrical function:
The suction motor relay connects on the Main control PCB
and the suction motor starts (if controlled by the unit).
Control voltage is connected to the microvac and flushing
valves.
Pneumatical function:
The microvac valve enables the suction closing mem-
brane to connect the vacuum to the upper chamber (and
to the suction hose). The intermediate membrane is
forced to expand by the vacuum in the upper chamber.
The pressure balancing membrane over the intermediate
membrane moves upwards and blocks the opening
between the upper chamber and the lower chamber.
Saliva and water begins to gather into the upper chamber.
Microvac valve
Suction closing
membrane
Intermediate
membrane
Pressure balancing
membrane
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......