Chapter F - ADJUSTMENTS
Planmeca Compact i F-9
MECHANICAL ADJUSTMENTS
Technical Manual
1.4 Adjusting angle of the OP delivery arm instrument console
a)
Remove the lifting arm cover.
b)
Remove the ø8 DIN 471 locking ring and pull out the lower joint pin (short).
c)
Adjust the angle of the instrument console by turning the lower joint adjustment screw with a
15mm set wrench. The angle is preset at the factory to level with the rotating arm.
d)
Assemble the lower joint in reverse order.
S
C
D
A
B
8
8
8
8
The angle is preset at the factory
to level with the rotating arm.
Lifting arm cover
Lower joint
pin (short)
Locking ring
ø8 DIN 471
Lower joint
adjustment screw