Chapter E - PREVENTIVE MAINTENANCE
E-6 Planmeca Compact i
WATER MANAGEMENT SYSTEM (WMS)
Technical Manual
11 WATER MANAGEMENT SYSTEM (WMS)
Checking / adjusting the water flow
•
Open the cuspidor door and switch on the pump (bottle switch in ON position). Close the
door.
•
Enter the service mode 127.
•
Enter the SET-mode by pressing and holding the “S”-key until the yellow indicator light
turns on.
•
Run water from the glass fill tube until the separation tank is empty (you will hear when the
tank is empty). Note, that the water will not flow if you do not have glass in the holder.
•
Open the cuspidor door. Detach the pump water tube from the container quick connector
and place it into the
empty
container as shown on the figure below.
•
The water will flow for one minute when you press the B-key briefly. The water flow can be
stopped by pressing the A-key, if needed.
•
Press the B-key briefly and let the water run for one minute. Pour the water into a glass
measure or measure the water level on the container: 100 mm water level is equivalent to
330 ml, 120 mm = 400 ml and 140 mm = 500 ml.
•
The water flow must be:
Metasys typ1 and Compact, Microvac, no suction: 330 - 400 ml / min.
Dürr and Dürr VSA systems: 400 - 500 ml /min.
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......