Chapter E - PREVENTIVE MAINTENANCE
E-2 Planmeca Compact i
INSTRUMENT CONSOLE
Technical Manual
3 INSTRUMENT CONSOLE
Oil collector
•
Clean the collector.
Instrument arms
•
Condition of the arms. Replace if needed.
Instrument hoses
•
Condition of hoses. Replace if needed.
•
Condition of quick connectors.
•
Condition of rubber gasket. Replace if needed.
•
Check for leaks.
Control panel
•
All the keys are working.
4 INSTRUMENTS
Syringe
•
Tip is clean.
Turbine
•
Drive air pressure. Service mode 30.
•
Bulb voltage.
•
O-rings. Change automatically twice a year.
•
Lubricate.
Micromotor
•
Bulb voltage.
•
O-rings. Change automatically twice a year.
•
Lubricate.
Polymerization light
•
Bulb voltage.
Recommended voltages are 8.90 Vdc (Lysta), 15.4 Vdc
(Satelec), 10.0 Vdc (Demetron, PowerLysta).