Foot control related error messages
ERROR
CODE
EXPLANATION
E51.5
Foot control's chair control knob was not in home position
when dental unit was switched on.
Restart the unit without touching the foot control.
E51.6
Foot control left knob is either stuck or active at power-up or
after pedal use.
E51.7
Foot control right knob is either stuck or active at power-up or
after pedal use.
E52
Foot control is of new type, update MCB software.
MCB software must be updated.
E52.1
Incompatible wireless foot control receiver software.
E52.2
Incompatible wireless foot control software.
E52.3
Incompatible foot control software.
E53
Foot control data error (cable short circuit).
E54.0
Foot control error.
E54.1
Foot control EEPROM write failure.
E54.2
Foot control EEPROM read failure.
E54.3
Pedal too far from sensors, channel A.
E54.4
Pedal too far from sensors, channel B.
E54.5
Sensor plate is tilted too much to either side, checksum error.
E54.6
Sensor plate is too close to PCB, channel A.
E54.7
Sensor plate is too close to PCB, channel B.
E54.8
Foot control re-trigger error.
E54.9
Foot control calibration error.
E55
Wireless foot control connection lost.
31.3.8 Operating light related error messages
Operating light related error messages
ERROR
CODE
EXPLANATION
E56
Operating light erroneously connected or short circuit in push
button.
E57
Incompatible SingLED software.
E58
Operating light bulb burned out.
Replace bulb.
31 Help and error messages
272 Planmeca Compact i
User's manual
Summary of Contents for Compact i Touch v2
Page 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Page 10: ...Table of contents Planmeca Compact i User s manual ...
Page 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Page 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Page 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Page 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Page 313: ......