23.7 Turbine
The turbines supplied by Planmeca have a built-in backflow prevention
system, which protects the water used in the instrument from contamination.
CAUTION
A power cut will shut down the software-controlled backflow prevention
system. If you are using a turbine without a built-in backflow prevention
system, contaminated water can enter into the turbine and turbine hose in
the event of a power cut.