![PIETRO FIORENTINI M-LOG Plus Technical Manual Download Page 19](http://html.mh-extra.com/html/pietro-fiorentini/m-log-plus/m-log-plus_technical-manual_1555803019.webp)
MT334-I
Rev. 0.3 08/04/2022
19 - 22
Document classified as: 2 - Company Confidential
11.1
I/O characteristic parameters
Digital Inputs 1-8
Parameter
Uo
Io
Po
Ui
Ii
Pi
Co
Lo
Ci
Li
Value (Ver. A)
6 V
6,5mA
10mW
6 V
N.A.
N.A.
10μF
47μH
Nota 1
5μH
Value (Ver. B)
6 V
6,5mA
10mW
6 V
N.A.
N.A.
10μF
47μH
Nota 1
2,5μH
Value (Ver. T)
6 V
6,5mA
10mW
6 V
N.A.
N.A.
10μF
47μH
Nota 1 50nH
Digital Inputs 9-10
Value (Ver. A)
6 V
6,5mA
10mW
15 V
N.A.
N.A.
10μF
47μH
Nota 1
5μH
Value (Ver. B)
6 V
6,5mA
10mW
15 V
N.A.
N.A.
10μF
47μH
Nota 1 2,5µH
Value (Ver. T)
6 V
6,5mA
10mW
15 V
N.A.
N.A.
10μF
47μH
Nota 1 50nH
Digital Outputs
Parameter
Uo
Io
Po
Ui
Ii
Pi
Co
Lo
Ci
Li
Value (Ver. A)
6 V
6,5mA
10mW
15V
N.A.
N.A.
10μF
47μH
Nota 1
5μH
Value (Ver. B)
6 V
6,5mA
10mW
15V
N.A.
N.A.
10μF
47μH
Nota 1
2,5μH
Value (Ver. T)
6 V
6,5mA
10mW
15V
N.A.
N.A.
10μF
47μH
Nota 1 50nH
Inputs for Analog pressure sensor (1)
Parameter
Uo
Io
Po
Ui
Ii
Pi
Co
Lo
Ci
Li
Value
6 V
222mA
333mW
N.A.
N.A.
N.A.
10μF
47μH
N.A.
N.A.
Auxiliary Analog inputs 0-5V
Parameter
Uo
Io
Po
Ui
Ii
Pi
Co
Lo
Ci
Li
Value (Ver. A)
6 V
1mA
6mW
6 V
N.A.
N.A.
10μF
47μH
450μF
5μH
Value (Ver. B)
6 V
1mA
6mW
6 V
N.A.
N.A.
10μF
47μH
85μF
2,5μH
Value (Ver. T)
6 V
1mA
6mW
6 V
N.A.
N.A.
10μF
47μH
450μF
50nH
Input for Temperature sensor
Parameter
Uo
Io
Po
Ui
Ii
Pi
Co
Lo
Ci
Li
Value
6 V
26mA
39mW
N.A.
N.A.
N.A.
10μF
47μH
N.A.
N.A.
RS485 Serial Expansion Line
Parameter
Uo
Io
Po
Ui
Ii
Pi
Co
Lo
Ci
Li
Value (Ver. A)
6 V
6,5mA
10mW
6V
N.A.
N.A.
180 μF
200μH
450μF
5μH
Value (Ver. B)
6 V
6,5mA
10mW
6V
N.A.
N.A.
500 μF
200μH
85μF
2,5μH
Value (Ver. T)
6 V
6,5mA
10mW
6V
N.A.
N.A.
180 μF
200μH
450μF
50nH
(1)
The following transducer models can be connected to these channels:
STS
TD GAS, STS
TM EX
GEMS 563966, GEMS 564280
TECLAB GSE-03, TECLAB TL01, TECLAB GSE/03/1
FAST GSE-03, FAST TL01, FAST
GSE/03/1
Or any ATEX certified transducer with characteristic parameters compatible with the values shown in the table.