PicoVNA 106 6 GHz Vector Network Analyzer
49
© 2017 Pico Technology
pv106ug r1
Measured response of 50cm shorted cable using tenth-order Kaiser-Bessel window.
A more complicated example:
The trace below shows the response of a 30cm, 50 Ω line followed by 30cm of 25Ω line terminated in a
short circuit. The window is third order Kaiser–Bessel. The trace shows the multiple reflections from the
discontinuity of impedance at the connection between the lines and the short circuit termination.
Measured response of multiple reflections. See text for details.
The last example is the same 50 Ω / 25 Ω cable combination but this time the termination is an open circuit.
The response is also displayed over a longer period.
Measured response of multiple reflections. See text for details.