PicoSource PG900 software
26
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
4.3
Tunnel diode pulse head outputs
TD Head 1 Output (left) and TD 2 Head Output (right) menus
The
TD Head 1 Output
and
TD Head 2 Output
menus control the outputs of two TD
heads. These heads may be connected to the
DRIVE 1
or
DRIVE 2
SMA connectors
located on the front panel of the PicoSource PG912 and PG914. Each drive output
provides the head the necessary TD current and drive signal.
Each drive output senses the polarity of the TD head connected to the drive output.
Thereby it is possible to connect as positive or as negative TD head to any
DRIVE
output. The drive output will provide a positive current and positive drive output for
the positive TD head, or a negative current and negative drive output for the negative
TD head, depending on which is detected. The recognized polarity of the TD head
appears in the menu area under
Pos TD Head
or
Neg TD Head
.
Pulse generators based on TD technology may include the following configurations:
·
Single-channel pulse generator with TD pulse head having any polarity.
·
Dual-channel pulse generator with two TD pulse heads having opposite polarity.
·
Dual-channel pulse generator with two TD pulse heads having the same polarity. In
this case you should request an additional TD head.