PicoSource PG900 Series User's Guide
19
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
3.4
External trigger operation
To operate the external trigger follow the setup instructions below.
1.
For external trigger operation connect external trigger source to the
TRIGGER
INPUT
connector.
2.
Select the
Ext Trigger
from
Period Source
. The
External Trigger
menu will
then open:
3.
Under
Slope
, select
Positive
or
Negative
trigger slope.
4.
Adjust the
Level
control from –1 V to 1 V as necessary for stable operation.
5.
Set necessary
Holdoff
control
from 1 µs to 1 s. Trigger holdoff helps to stabilize
triggering of the generator. When you adjust the
Holdoff
control, the amount of
time that the generator waits before re-arming the trigger circuitry also changes.
This allows you to provide stable trigger up to 1 GHz.
Summary of Contents for PicoSource PG900 Series
Page 2: ......