10 Technical Data
50
Version: 1.0.1
MP134E
Q-522 Miniature Stage
10.2 Operating Time
The operating time and the operating frequency in step mode affect the lifetime of the
stage. In order to prevent overheating and high wear, the operating time with given
operating frequency and 100 % duty cycle should not exceed the values given in the
following table.
Operating frequency in Hz
1
Operating time
2
/ 48 V, ±0.2 A
20000
10 s (max.)
10000
20 s (max.)
5000
60 s (max.)
≤
1000
120 s (max.)
1
For the relevant parameters see "Starting Up the Stage" (p. 36) and the user manual
of the electronics used.
2
With 100 % duty cycle without heat dissipation