![PI PIMag V-408 User Manual Download Page 14](http://html1.mh-extra.com/html/pi/pimag-v-408/pimag-v-408_user-manual_1554464014.webp)
6
Installation
6.1
Mounting the V-408
The V-408 can be mounted onto a surface or an optical table via the V-500.AP1 adapter.
The V-408 can be mounted onto a surface from below or from above.
Overview
Figure 5: Holes for mounting onto a surface or onto the V-500.AP1 adapter
The arrows point to the following holes in the underneath of the V-408:
Black arrows
(align):
Locating holes Ø 3 mm H7, depth 3.5 mm
White arrows
(mounting from above):
Countersunk holes M2.5 (Ø 2.9 mm / Ø 5.5 mm), depth 2.9
mm
Gray arrows
(mounting from below):
Threaded holes M3, depth 6 mm
Tools and Accessories
■
Optional: 2 locating pins of suitable length with Ø3 mm h8
■
For mounting from above: 4 M2.5 screws of suitable length
■
For mounting from below: 4 M3 screws of suitable length
■
Suitable screwdriver
Requirements
✓
You have read and understood the
general safety instructions (p. 8)
✓
You have provided a suitable underlying surface with the holes necessary for the screws
and, if required, locating pins. For the position and depth required for the holes, see
■
The surface flatness is
≤
20 μm.
■
For applications with large temperature fluctuations: The surface should have the
same thermal expansion properties as the V-408 (e.g., underlying surface made of
aluminum).
6 INSTALLATION
V408M0001EN
‒
2/17/2020
14