9
Troubleshooting
No or limited motion
Cable not connected correctly
►
Check the cable connections.
Excessive load
►
Do not exceed the maximum push/pull capacity
according to the
Piezo actuator is depolarized due to over-
heating
►
Contact our
.
Defective actuator
►
If possible, replace the defective actuator with
another actuator and test the new combination.
The piezo actuator starts oscillating or positions inaccurately
Operating parameters wrongly set because
for example, the load was changed
►
Switch off the electronics immediately.
►
Check the settings of the operating parameters
(see the user manual for the electronics).
►
If necessary, correct the settings for the operat‐
ing parameters.
Operating with an excessively high frequen-
cy
►
Operate the piezo actuator at a maximum of one
third of the resonant frequency (see
for resonant frequency of the un‐
loaded piezo actuator, see
mum Operating Frequency of the Loaded Piezo
for resonant frequency of the
loaded piezo actuator).
9 TROUBLESHOOTING
P840M0001EN
‒
1/29/2021
25
M O T I O N | P O S I T I O N I N G