5 Installation
P-88x.x5 Piezo Actuator
PZ266E
Version: 1.1.0
21
5.4
Connecting the P-88x.x5 to the Electronics
The P-88x.x5 piezo actuator is connected to a LEMO socket, a terminal, or soldering pins,
depending on the electronics.
This section describes how the P-88x.x5 piezo actuator is connected with the P-890.xx coaxial
cable to electronics with a LEMO coaxial socket.
For connection to the electronics with other LEMO connectors, contact our customer
service department (p. 37).
If you use a self-made connecting cable instead of the P-890.xx coaxial cable, pay
attention to the relevant standards as well as the assembly information of the
manufacturer of the connector used.
For connection to terminals or solder pins, refer to the manual of the respective
electronics used (p. 2).
INFORMATION
The P-88x.x5 has color-coded stranded wires:
Red stranded wire: Voltage connection (+)
Black stranded wire: Ground (-)
Figure 9: Connection of the P-88x.x5 piezo actuator to the P-890.xx coaxial cable
+
On the piezo actuator (left): Red stranded wire for voltage connection
On the LEMO connector (right): Inner contact for voltage connection
-
On the piezo actuator (left): Black stranded wire for ground
On the LEMO connector (right): Connector housing
S
Cable shield
LEMO
LEMO FFS.00.250.CTCE24 coaxial connector
Requirements
You have read and understood the general notes on installation (p. 15).
You have read and understood the user manual of the electronics used.
If the P-88x.x5 is not short-circuited: The P-88x.x5 is discharged (p. 30).
The electronics are
switched off
.
The electronics have a suitable LEMO socket for the connector on the P-890.xx coaxial
cable.