The value corresponds to one third of the resonant frequency of the unloaded piezo
actuator. For further restrictions, see
"Overview of Limiting Factors" (p. 16)
■
Maximum operating frequency without load, considering thermal aspects, column B:
In order to prevent the maximum permissible operating temperature from being
exceeded, the operating frequency of the unloaded, uncooled piezo actuator must not
exceed the specified frequency when the operating voltage is 100 V peak-to-peak. In the
case of smaller amplitudes of the operating voltage and/or the use of cooling measures,
higher operating frequencies are possible. For further restrictions, see
■
Maximum power consumption:
The value corresponds to the power consumption of the unloaded, uncooled piezo
actuator that is operated at a voltage of 100 V peak-to-peak with the operating frequency
from column B of this table.
Piezo actua-
tor*
Maximum
operating
voltage
range
Maximum operating frequency
without load
Maximum pow-
er consumption
A:
Without consid-
ering thermal
aspects
B:
Considering
thermal aspects
Considering
thermal aspects
P-84x.10
0 to 100 V
5333 Hz
55 Hz
5.7 W
P-84x.20
0 to 100 V
4000 Hz
48 Hz
11.4 W
P-84x.30
0 to 100 V
3000 Hz
45 Hz
17.1 W
P-84x.40
0 to 100 V
2500 Hz
42 Hz
22.8 W
P-84x.60
0 to 100 V
1833 Hz
39 Hz
34.2 W
* The letter x in the product number of the piezo actuator stands for "4" or "5".
12.3
Ambient Conditions and Classifications
The following ambient conditions and classifications for the P-844 must be observed:
12 TECHNICAL DATA
P844M0001EN
‒
11/13/2018
27
M O T I O N | P O S I T I O N I N G