Mounting the P-725 using the mounting holes in the base body
1.
Mount the thread adapter for the objective side to the P-725 (depending on model):
−
P-721.xxQ thread adapter set:
Screw the thread adapter for the objective side into the aperture of the P-725 from
below. For details, refer to "P-721.xxQ Thread Adapter Set (for P-725.xxD/.xxL and
P-721.xxQ/.SL2)" (p. 14).
−
P-721.xxA thread adapter set:
Screw the thread adapter for the objective side into the aperture of the P-725 from
below. For details, refer to "P-721.xxA Thread Adapter Set (for P-721.xxA and P-
−
P-725.xxL thread adapter:
Clamp the thread adapter in the aperture of the P-725. For details, refer to "P-
725.xxM and P-725.xxL Thread Adapters (for P-725.1CDE2)" (p. 22).
2.
Optional: Align the P-725 to the microscope. Use suitable locating pins and the locating
holes in the base body of the P-725.
3.
Fix the P-725 to the microscope. Use four suitable M2.5 screws and the mounting holes
in the base body of the P-725 for fixing.
4.
Screw the objective from below into the thread adapter on the objective side of the
PIFOC positioner. Do not exceed the maximum torque of 0.6 Nm.
Connecting the P-725 to the Electronics
Requirements
You have read and understood the safety precautions (p. 6).
You have read and understood the user manual for the electronics.
The electronics are switched off, i.e.,
not
connected to the power supply.
Connecting the P-725 to the Electronics
Connect the plugs of the P-725 to the corresponding sockets on the electronics (refer to
the user manual for the electronics).