6 Start-Up and Operation
P-56x Nanopositioners
PZ248E
Version: 1.0.0
29
INFORMATION
Systems are calibrated at the factory to achieve optimum positioning accuracy.
Replacing the system components will cause a loss in positioning accuracy in the
following cases:
ID chip (p. 14) of the stage does not contain any calibration data
Stage is connected via LEMO connectors to the controller (no ID chip lines
present)
When connecting the stage, observe the assignment of the stage axes to the
controller channels, which is given by the calibration label of the controller.
If the positioning accuracy is reduced after the P-56x, the controller, an extension
cable or an adapter cable has been replaced:
Perform a recalibration of the axis displacement (see controller manual) or
contact our customer service department (p. 37).
INFORMATION
Sound and vibration (e.g. footfall, impacts) can be transmitted to the stage and can
affect its performance with regard to position stability.
Avoid transmitting sound and vibration while the stage is being operated.
6.2 Operating the P-56x
Prerequisite
If a P-56x.3CL stage is to be operated with a controller with LEMO sockets:
The E71000211 adapter cable is connected to the stage (p. 25).
Operating the P-56x
Follow the instructions in the manual of the used controller for start-up and
operation of the P-56x.