5 Installation
N565 NEXACT® Linear Stage
MP148E
Version: 1.4.0
15
NOTICE
Damage due to improper mounting!
Improper mounting of the N-565 or incorrectly mounted parts can damage the N-565.
Only use the holes or threads intended for the purpose of fixing the N-565 and loads.
Install the N-565 so that the platform and all parts attached to it can move freely within the
entire travel range.
NOTICE
Warping of the N565 when mounted on uneven surfaces!
The N-565 could warp if mounted on an uneven surface. Warping reduces the accuracy.
Mount the N-565 onto an even surface. The recommended flatness of the surface is
≤10
µm.
For applications with large temperature changes:
Mount the N-565 only onto surfaces that have the same or similar thermal expansion
properties as the N-565.
NOTICE
Damage due to unsuitable cables!
Unsuitable cables can damage the N-565 and the electronics.
Use cables provided by PI only to connect the N-565 to the electronics.
INFORMATION
For optimum repeatability, all components must be connected firmly together.
5.2
Mounting the N565 onto a Surface
Options for mounting the N565 onto an underlying surface
INFORMATION
The mounting holes in the base body of the positioner are only accessible from above when
the platform is in the reference position (delivery state). If necessary:
Start the positioner (p. 23) and do a reference move to the reference switch (refer to the
user manual for the controller).