User Manual
F712T0002, valid for F-712.MA1 and F-712.MA2
Mma, BRo, 12/6/2019
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 36 / 45
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Requirements for the Optical Power Meter Used
Value
Unit
Output signal
Analog output, ideally converted from linear to logarithmic
Output voltage range, max.
-5 to 5
V
Bandwidth, min.
1
kHz
Noise level, max.
-60
dBm
Details of E-712K236 / E-712K252 Controllers
Usable analog input channels:
E-712K236, E-712K252
Connection
In 1
to
In 4
sockets of the E-711.IA4 analog interface module
Channel identifiers
Four channels. For possible identifiers, see “Fast Alignment Input Channels Group”
in the E712T0016 user manual.
Input voltage range
±10 V differential
Resolution ADC
18 bit
Bandwidth
25 kHz
Input impedance
150 kohm
Connector
LEMO EPG.00.302.NLN
For detailed specifications of the E-712 controllers, see the controller documentation listed on
p. 5.