background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

MIC-200 Series / 1720

 

 

 
The  total  magnification  can  be  easily  calculated  by  multiply-
ing the magnification of the eyepiece by that of the respective 
objective, as shown in the table: 
 

Ocular 

Objective 

Magnification 

10x 

4x 

40x 

10x 

10x 

100x 

10x 

40x 

400x 

10x 

100x 

1000x 

 
In  MIC-200  series  microscopes,  the  objectives  are  already 
attached to the nosepiece. 
 
Check  that  all  the  objectives  are  firmly  attached.  To  do  this, 
grasp the knurled rings of the objectives with your fingers and 
turn them to the right. Do not use a tool to tighten them. The 
entire nosepiece with the 4 objectives can be turned by hand. 
When you rotate the nosepiece, you will notice that the objec-
tives click into place. They do so in the correct position rela-
tive to the optical axis of the microscope. 
 

Object table with  mechanical stage 

The microscope slide is placed in the specimen holder of the 
mechanical  stage  and  can  be  moved  in  X  and  Y  direction. 
The  stage  has  a  working  area  of  141  x  131  mm,  the  travel 
range is 78 x 53 mm. 
 

Coarse and fine adjustment 

The focusing mechanisms for coarse and fine adjustment are 
located  on  one  axis  (coaxial).  On  the  fine  adjustment  knobs 
there is a graduation. This can be used to measure the depth 
of a specimen. 
When the stage automatically slides down after using the mi-
croscope  for  some  time,  turn  the  adjustment  ring  on  the  in-
side  of  the  coarse  and  fine  adjustment  knobs  slightly  in  the 
direction  of  the  arrow.  The  coarse  adjustment  knob  is  tight-
ened to prevent the stage from sliding down. 

 
Abbe condenser with iris diaphragm 

Below the object stage is an Abbe condenser N.A. 1.25. The 
condenser  can  be  adjusted  in  height  by  turning.  This  allows 
light  to  be  focused  on  the  specimen  for  contrast  enhance-
ment. The condenser is already pre-centered. 
The iris diaphragm with filter holder is located under the con-
denser.  The  light  intensity  can  be  adjusted  by  changing  the 
movable opening. 

 
Illumination 

The  power-saving  illumination  is  the  reason  why  the  micro-
scopes of the MIC-222 model do not need to be connected to 
the mains all the time for operation, but can also be operated 
independently  of  the  mains,  as  they  contain  a  rechargeable 
Ni-MH  battery.  The  illumination  is  a  modern  LED  lamp  with 
the following specifications: 

  LED: 3W, brightness adjustable 

  Power supply: 100V-230V 

 
C-mount adapter for the trinocular version MIC-231 

A C-mount adapter is part of the scope of delivery of the MIC-
231 so that you can use each microscope camera which has 
a C-mount thread. 

5  WORKING WITH THE MICROSCOPE 

Please  follow  the  following  instructions  to  achieve  the  best 
results: 
 

Adjustment  of  interpupillary  distance  and  diopter  dis-
tance 

1. Adjusting the interpupillary distance 
While looking through the eyepieces, adjust them until the left 
and right fields of view coincide completely. 

 

2. Adjusting the diopter 
a) While looking through the right eyepiece, rotate the 
coarse/fine focusing adjustment knobs to bring the specimen 
into focus. 
b) Look through the left eyepiece and rotate only the diopter 
adjustment ring on the left eyepiece sleeve to bring the spec-
imen into focus. 

 
Adjusting the lighting

 

For  optimal  contrast  and  resolution,  please  follow  these 
steps: 

  Place  a  micro  slide  on  the  stage  and  focus  with  the  4x 

lens. The iris diaphragm is open. 

  Turn the condenser to the highest position. 

  Close the iris diaphragm until it is just visible at the edge 

of the image field. 

  The microscope is now optimized for use with the 4x ob-

jective. Repeat this procedure with each additional magni-
fication to achieve the best balance between contrast and 
resolution. 
 

Attention: the maximum light intensity of the microscope 
can  damage  your  eyes  if  you  use  it  with  the  4x  and  10x 
objectives! 
 

If you swap micro slides, start again with the 4x objective. 
It is recommended to use slides of 1.0 to 1.2 mm thickness in 
combination  with  cover  glasses  of  0.13  mm  or  0.17  mm 
thickness. 

 
Battery 

The  MIC-222  microscope  contains  a  3.6V  Ni-MH  battery 
pack. 
A charging indicator is located on the side of the microscope. 
It is red when connected to the power supply and green when 
the battery is full. If the red indicator flickers,

 

the battery is not 

properly connected to the circuit board. 
The  microscope's  illumination  can  be  powered  by  a  battery 
pack for 4 to 8 hours without an external power supply. 

 
Using the S100x oil immersion objective 

Please follow the following recommendations for using S100x 
N.A. 1.25 oil immersion lens: 
 

Reviews: