5.7 Possible sources of error. 

 

1.  If  the  electrode  membrane  is  not  mounted  tightly, 

measuring solution enters the inside of the electrode 
and  falsifies  the  measured  values.  Replace  the 
conductive liquid as described in 5.2. 

2.  Sediments  settle  on  the  electrode  membrane  and 

block the membrane's function. Replace the memb-
rane and ensure careful cleaning and storage. 

3.  The service  life  of  an  ammonia  electrode  is approx. 

1  year,  regardless  of  use  or  storage.  Older  elec-
trodes  may  have  an  adverse  effect  on  measure-
ments.  Therefore,  replace  the  electrode  after  ap-
prox. 1 year. 

 
 
5.8 Replacing the battery 
 
Remove the battery 

 

Open  the  sensor  by  turning  the  screw 
cap  on  the  back  of  the  sensor  counter-
clockwise, e.g. with a coin. 
 
 
 
 
 

  

 

 Fig. 8 

 
Lever  the  battery  sensitively, 
e.g.  with  the  help  of  a  small 
screwdriver  or  a  small  pair  of 
scissors,  out  of  its  socket.  In-
sert  the  screwdriver  as  shown 
in Fig. 9. 

 

 

    

 

 

 

 

       

 

 

 

 

 

 

Fig. 9 

Insert new battery 

 
Slide  the  battery  under  the  golden  metal  nose  (Fig.10-1). 
Make  sure  that  the  battery  is  completely  under  the  metal 
nose and completely pushed to the upper edge. 

Push  the  battery  into  the  socket 
by  pressing  lightly  on  the  oppo-
site side.  
The  battery  slips  under  the  two 
plastic  lugs  (Fig.  10-2),  which  is 
also noticed by a short "click". 
 
 

              
 

    

 

 

    

Fig. 10 

 
 

 

Before  closing,  make  sure  that 
the seal in the lid is not bent and 
lies neatly on the edge of the lid. 
Then  tighten  the  cover  clock-
wise. 

 

 
 
  Fig. 11 

6  TECHNICAL DATA  

Operating temperature range: 5 - 40°C 
Relative humidity < 80%. 
 

 

Measuring range 
Resolution 
Accuracy* 

0.

9… 1800 mg/l 

0.5 mg/l 

±10% 

Max. data transfer rate 
Battery type 
Max. wireless range (open field)  
Dimensions (WxHxD) 

10 Hz 

CR2032 

30 m 

90 x 44 x 23 mm 

Weight (incl. electrode) 
 
*After calibration 

97 g 

7  SCOPE OF DELIVERY 

The scope of delivery includes: 

 

Cobra SMARTsense Ammonium Ion 

12913-00 

 

Ammonium Ion-selective electrode PNH

3

-2-004 

 

Instruction manual 

8  ACCESSORIES 

The following accessories are available: 

 

Button cell CR2032, 3 V 

   07922-15 

 

Cobra SMARTlink 

   12999-99 

 

Ammonium Ion-selective electrode 
PNH

3

-2-004 

   12913-10 

 

Free measureApp available from supplier portals 

iOS 

 

Android 

 

Windows 

 

 

 

9    CONFORMITY 

 

 

 
 

 
 
 

 

PHYWE Systeme GmbH & Co.KG hereby 
declares that the radio system type 12913-
00 complies with the 2014/53/EU directive. 
The complete text of the EC Declaration of 
Conformity is available at the following In-
ternet address:  

www.phywe.com/en/ec-declaration

 

 
 

Summary of Contents for Cobra SMARTsense Ammonium Ion 12913-00

Page 1: ...RIES 9 CONFORMITY 10 DISPOSAL 1 SAFETY PRECAUTIONS Carefully read these operating instructions completely before operating this instrument This is necessary to avoid damage to it as well as for user safety Only use the instrument for the purpose for which it was designed Only use the instrument in dry rooms in which there is no risk of explosion Protect the instrument from dust moisture and vapour...

Page 2: ...urs unplug it and plug it back in again for a reset 5 HANDLING This section describes the start up of the sensor and the re cording of measurement data Please read this section thor oughly in order to avoid failures or operating errors 5 1 Start up Switch the sensor on by pressing the on button for more than 3 seconds The Bluetooth LED lights up red Star the soft ware and select the sensor There i...

Page 3: ...0 ml H20 deionised 10 ml 1 M NH4Cl 5 4 Calibration Perform a 2 point calibration using measureAPP Use 160 mg l for the 1st measuring point and 1600 mg l for the 2nd measuring point The calibration solutions are prepared from the stock solutions Use a magnetic stirrer with uniform stir ring speed for uniform mixing Recipe for calibration solutions To prepare 1 litre of calibration solution you will...

Page 4: ...hed to the upper edge Push the battery into the socket by pressing lightly on the oppo site side The battery slips under the two plastic lugs Fig 10 2 which is also noticed by a short click Fig 10 Before closing make sure that the seal in the lid is not bent and lies neatly on the edge of the lid Then tighten the cover clock wise Fig 11 6 TECHNICAL DATA Operating temperature range 5 40 C Relative ...

Page 5: ... returned to the local recycling sta tions PHYWE Systeme GmbH Co KG Abteilung Kundendienst Robert Bosch Breite 10 D 37079 Göttingen Telefon 49 0 551 604 274 Fax 49 0 551 604 246 Do not dispose of this product with normal household waste If this unit needs to be disposed of please return it to the address that is stated below for proper disposal ...

Reviews: