background image

 

 

 

 

 

 

10 

www.phywe.com, © All rights reserved 

 

12620-09 and 12620-10 / 2715

 

After  an unsuccessful calibration,  the  values can be  reset to 
the  factory  settings  by  way  of  "Reset"  (4.3).  An  "attention" 
symbol  will  be  displayed  requiring  the  user  to  confirm  the 
entry.  

 

Note: 

If  the  Cobra4  Sensor-Unit  Condu  (12632-00)  is 

used,  the  menu  item  under  4.3  is  the  item 

Cell  constant. 

Here, the cell constant of the conductivity sensor that is used 
can be entered and saved.  

 

5.4.7

 

Setup/Date/Time (5.1) 

The date and time can be selected by way of the navigation 
keys 

▲▼ and they can be adjusted with ◄►. 

 

5.4.8

 

Setup/Battery status (5.2)

 

The  current  consumption  of  the  activated  Cobra4  Mobile-
Links  leads  to  a  continuous  decrease  of  the  supply  voltage. 
This menu can be used to enquire the battery status. In addi-
tion,  the  total capacity  of  the battery,  the  charging  time,  and 
the remaining capacity are displayed (Fig. 24). 

If  the  capacity  of  the  battery  falls  below  a  critical  threshold 
(10

 %  of  the  remaining  battery  capacity)  the  battery  symbol 

will  be  displayed  in  red  in the  header  of the display.  Prior  to 
the  automatic  shut-down  of  the  device,  any  running  meas-
urements will be stored automatically. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.4.9

 

Setup/Device (5. 3)

 

Selection  of  the  device  properties  by  way  of  the  arrow 
keys 

▼▲ and confirmation by way of the confirmation key 

The following settings can be realised by way of ◄►:  

-

 

Language of the device menu. You can choose between 

German 

, English 

, and Russian 

 . 

-

 

Sound: The sound of the keystrokes of the device can be 
switched on 

 or off 

-

 

GPS:  (option)  In  order  to  localise  the  device,  a  satellite 
connection can be set up. This function is switched off as 
standard.  After  it  has  been  switched  on  by  way  of  the 

switch,  the  GPS  symbol 

  will  be  displayed  in  the 

header. 

-

 

The  automatic  rotation  of  the  device  display  can  be 
switched on 

 and off 

Use the confirmation key 

 in order to confirm a new setting. 

 

5.4.10

 

Setup/Information (5.4)

 

Here,  you  can  find  general  information  concerning  your  de-
vice, e.g. the number of the firmware version, the number of 
devices, and the last update of the Mobile-Link.  

 

5.4.11

 

Setup/Large Display (5.5) 

Here,  the  settings  options  for  communication  with  up  to  two 
large-scale  display  units  (07152-93)  can  be  selected  if  a 
Display-Connect (12623-00) is attached. 

 

The  uppermost  channel  of  the  measurement  display  will  be 
transferred automatically to the first large-scale display unit. If 
a  second  large-scale  display  unit  is  connected,  the  second 
measuring channel that is to be displayed can be selected for 
five radio channels. 

 

5.4.12

 

Utilities/Stopwatch (6.1) 

This feature enables you to record a time interval in the range 
of a hundredth of a second. 

After  the  start  of  the  measurement  with 

,  it  is  possible  to 

pause the measurement with ► and to resume it afterwards. 
For every pause, one measurement value will be saved to a 
list.  Up  to  three  measuring  points  can  be  displayed  simulta-
neously.  Then,  the  first  one  will  be  overwritten.  Pressing 

 

again  stops  the  measurement  and  reset  the  stopwatch  to 
00:00:00. 

Pressing 

►  while  there  is  no  running  measurement  will  de-

lete all of the measurement values. 

The  various  activities  are  visualised  in  the  footer  of  the  dis-
play. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 25: Display  of  the  "Stopwatch"  utility  of  the  Cobra4  Mo-

bile-Link 2 after the second measurement 

Fig. 24: Display of the submenu "Battery status" for the Co-

bra4 Mobile-Link 2 

The  setup  data  will  be  maintained  in  a  bat-
tery-buffered manner for up to 8 years. How-
ever,  this  requires  regular  recharging  of  the 
built-in battery. 

Battery status 
 
Menu

 

Stopwatch

 

10:20:67

 

00:01:38 

10:20:67 

00:00:00 

 

Start

 

 

 

 

Capacity   

1905  mAh 

Remaining

  

1543

 

mAh 

Rem. Time

  

191

 

min 

 

81 %

 

 

 

 

 

 

      

Start

 

Menu > Setup

 

Summary of Contents for 12620-09

Page 1: ...nsor Setup 1 2 5 4 3 Datalogger Setup 2 1 5 4 4 Datalogger Files 2 2 5 4 5 Zero 3 5 4 6 Calibration 4 5 4 7 Setup Date Time 5 1 5 4 8 Setup Battery status 5 2 5 4 10 Setup Information 5 4 5 4 11 Setup...

Page 2: ...supply unit 3 FUNCTIONAL AND OPERATING ELEMENTS 3 1 Operating elements on the top The Cobra4 Mobile Link 2 is equipped with various elements for the measurement data display operation and status indi...

Page 3: ...nity of the unit After a complete shut down of the unit it must be switched off and on again by way of the ON OFF switch operating ele ment If the unit does not react keep the ON OFF switch pressed fo...

Page 4: ...performed up to three times 5 3 Measurement data recording process A measurement can be started and stopped by pressing the confirmation key The measurement data recording proc ess is indicated by a...

Page 5: ...The file name consists of the date and a running counter YYMMDDxx mli xx 00 99 Ensure that the date and time are set in the Cobra4 Mobile Link This can be checked and if needed changed in the Setup m...

Page 6: ...Unit Conductivity 12632 00 It enables the automatic temperature compensation of a pH or conductivity measuring channel e g T 1 with regard to a temperature input or the manual input of a known temper...

Page 7: ...e following Cobra4 Sen sor Units Cobra4 Sensor Unit Radioactivity 12665 00 It is used to change the counter tube voltage in order to measure the counter tube characteristics Gate time This setting opt...

Page 8: ...imer Counter 12651 00 5 4 3 Datalogger Setup 2 1 The following changes are possible Selection of the single value measurement mode by way of a switch This ensures that only one value will be re corded...

Page 9: ...u item enables the measurement values of the measuring channels to be manually reset to zero To do so the switches of the desired channels are set from off to on by way of the arrow keys Fig 20 You ca...

Page 10: ...played in the header The automatic rotation of the device display can be switched on and off Use the confirmation key in order to confirm a new setting 5 4 10 Setup Information 5 4 Here you can find g...

Page 11: ...es minutes and seconds This menu item can only be used in combination with a Mo bile Link 2 GPS 12620 20 and GPS activated See also chapter 5 4 9 Coordinates will not be displayed unless satellites ha...

Page 12: ...f the battery decreases strongly it is possible to change the battery Please note that damage resulting from a battery change are not covered by the warranty You can send the Mobile Link to the repair...

Page 13: ...Unit Energy 12656 00 Cobra4 Sensor Unit Forceplate 12661 00 Cobra4 Sensor Unit Radioactivity 12665 00 Cobra4 Sensor Unit Sound Level 12669 00 Cobra4 Sensor Unit Weather air pressure humidity temperatu...

Reviews: