LIFEPAK 20/20e Defibrillator/Monitor
Modes of Operation
Manual Mode
Turning On the Device
in Manual Mode
Manual mode enables the user to determine when to deliver a shock.
To configure the device to turn on in manual mode (the default is AED mode):
1. Display the
menu and select
MANUAL MODE
.
2. Select
MANUAL ACCESS
in the
MANUAL MODE
submenu.
3. Select the
DIRECT
option.
The following table shows all the available power-on options.
If the device is placed in manual mode and then reset to AED mode by pressing
the
ANALYZE
button, there are no additional manual mode reprompts or
passcode requests until the device power has been cycled.
Mode/Response
Response Description
Manual/Direct
Turns on in manual mode; direct access between AED
and manual modes.
AED/Direct
Turns on in AED mode; direct access between AED
and manual modes.
AED/Confirmed
Turns on in AED mode; confirmation required to enter
manual mode.
AED/Passcode
Turns on in AED mode; passcode required to enter
manual mode.
(Continued on next page)
4-2