LIFEPAK 20/20e Defibrillator/Monitor Replacement
Procedures
Bottom Case
(continued)
A12 Printer Module
Installation
(continued)
7. If removed, press the fastener (225) into the printer shroud to secure the J45
flex connector.
8. Insert the printer bezel (83) ensuring that it is flush with the bottom case.
9. Connect the W14 Printer Flex Cable to the J1 connector on the printer. The
cable should lay flat against the rear of the printer.
10. Ensure that the W14 Printer Flex Cable lays between the two captured
screws.
11. Slide the A12 Printer into the printer shroud.
12. Tighten the two captured screws located in the A12 Printer and torque to
10 in-lb.
13. Install the paper roll in the A12 Printer and close the printer door.
14.
15.
16.
.
17.
18. Complete the LIFEPAK 20 Device PIP.
83
A12
9-120