LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Battery Pins / Power PCB Cable (W10) Replacement
336
8
Installing the Battery Pins/Power PCB Cable (W10)
To install the W10 cable:
If required, replace EMI shield as described in
EMI Shield Replacement (p. 305)
1.
Install the battery pins/power PCB (W10) cable (V1-
2.
Connect terminals P43 and P44 at Battery Well 1, and P45 and P46 at Battery Well 2 and
replace the four Kep nuts (
) (torque to 4.0 in-lb using 1/4” socket and drive
adapter). See
.
3.
Install the power PCB as described in
Installing the Power PCB (A03) (p. 257)
4.
If present, install the SpO2 PCB as described in
Installing the SpO2 PCB (A16) (p. 280)
.
5.
Connect the battery pins/power PCB cable (W10) connector locking tabs to the power PCB
at J11. V1 device will also install W10 at J13.
6.
Install the NIBP/CO2 Module as described in
Installing the NIBP (A21)/CO2 (A23) Modules
.
7.
Install the OEM PCB as described in
Installing the OEM PCB (A06) (p. 265)
8.
Install the energy storage capacitor as described in
Installing the Energy Storage Capacitor
9.
Install the biphasic module and inductive resistor as described in
Module (A22)/Inductive Resistor (A14) (p. 302)
10. Install the transfer relay assembly as described in
Installing the Transfer Relay Assembly
P11
P13
P46
P45
P44
P43
to Power PCB (A03)
to Battery Well 1
to Battery Well 2
Figure 8.111—(V1) Power
PCB cable connections. V2
cable, only P11 is used.