LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP Connector Replacement
307
8
Removing the NIBP Connector
Refer to
To remove the NIBP connector from the parameter bezel (rear case—refer to
Inside Rear Case Diagrams (p. 228)
NOTE:
These steps include all parameter bezel options. Your device may not have some of these options.
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
).
5.
Remove the fitting nut on the back side of the bezel to remove the connector (
).
6.
Remove the NIBP connector from the bezel.