LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
SpO2 PCB (A16) Replacement
276
8
7.
Disconnect the SpO2/OEM cable (W21) (
) from OEM PCB J26.
8.
Carefully disconnect the SpO2 connector flex cable
(W22) (
CAUTION
POSSIBLE COMPONENT DAMAGE
The SpO2 connector
flex cable can be damaged during disassembly. If SpO2
connector is damaged, replace the SpO2 flex cable using
the
MASIMO Sp02 Panel Mount Cable Repair Kit (MIN
.
To remove the SpO2 PCB from the rear case (see
Figure 9.21 on p. 413
):
(Continued) 13 steps, (Page 2 of 5)
W21 - SpO2/
OEM cable
J1
connection
Figure 8.68—SpO2 PCB connections