LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Speed Dial Assembly (W15) Replacement
218
8
Speed Dial Assembly (W15) Replacement
Removing the Speed Dial Assembly (W15)
To remove the speed dial assembly (
1.
Disassemble the case as described in
.
2.
Press the connector locking tab, and then disconnect the speed dial
assembly cable from the interface PCB at J33.
3.
From the outside of the front case, grasp the speed dial assembly
knob (
) and, with steady smooth force, pull the knob off
the speed dial assembly shaft. Use a gripping tool if necessary, taking
care to avoid any damage.
cable
connection
to J33
Figure 8.26—Speed dial cable
switch
knob from
front
Figure 8.26—Speed dial knob