LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Interface PCB (A05) Replacement
193
8
Removing the Interface PCB (A05)
Refer to
Interface PCB Replacement Figures (p. 190)
Inside Front Case Diagram (p. 189)
.
To remove the interface PCB (
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Disconnect the connectors (from top left, counterclockwise) as follows (see
~ J31 – Spread the connector locking tab and disconnect the main keypad interface PCB cable (W13,
~ J37 – Press the connector locking tab and disconnect the backlight interface PCB cable (W06,
).
~ J36 – Disconnect the LCD interface PCB cable (W18,
~ J32 – Press the connector locking tab and disconnect the printer control keypad interface PCB cable (W12,
~ J34 – Press the connector locking tab and disconnect the speaker interface PCB wire (W17,
).
~ J35 – Spread the connector locking tabs and eject the printer interface cable (W16,
~ J33 – Press the connector locking tab and disconnect the cable for the speed dial (W15,
3.
(If replacing the interface PCB) Disconnect the System/Interface cable as described in
System/Interface PCB Cable (W04)
4.
Remove the screw holding the ground strap. Discard the screw.
5.
Remove the eight remaining screws (
). Discard the screws.