LIFEPAK 15 Monitor/Defibrillator
Service Manual
Preventive Maintenance
Environmental Conditions
155
6
Environmental Conditions
Operating Conditions
Maintain the following operating temperatures when the device is in use:
•
32° to 113°F (0° to 45°C)
•
-4°F (-20°C) for 1 hour after storage at room temperature
•
140°F (60°C) for 1 hour after storage at room temperature
Maintain the following relative humidity when the device is in use:
•
5 to 95%, non-condensing
•
NIBP: 15 to 95%, non-condensing
Storage Conditions
When the device is not in use, store as follows:
•
-4° to 149°F (-20° to 65°C), except therapy electrodes and batteries
Maintain the following relative humidity when the device is in storage:
•
10 to 95%, non-condensing