LIFEPAK 15 Monitor/Defibrillator
Service Manual
Preventive Maintenance
Device User Test
147
6
Device User Test
Device User test
When you use the SPEED DIAL to navigate to Options/User Test, the device waits until the next self-test cycle is complete. The user
test performs the following tasks:
•
Performs a self-test
•
Charges the defibrillation capacitor to 10 joules, then dumps the charge
•
Stores the test results in the test log
•
Displays User Test results and prints report
If this operation does not pass, the Service LED illuminates and an error is written to the
.