Table 5.8—Initial Digit 3, Data Management Service Codes (DM)
Code
Service Code Description
Corrective
Action Code
3005
DM_ERROR_DATABASE_ERASE_ADJUST (Not able to write new lines for new oldest record; disables flash.)
3006
DM_ERROR_DATABASE_ERASE_FAILED (Erase block failed; disables flash; param = block requested.)
3007
DM_ERROR_DATABASE_ERASE_VERIFY (Verification of erased block failed; disabled flash; param = block.)
3008
DM_ERROR_DATABASE_FLASH_ERASE (Erase database failed; disables flash; param = block # of failure.)
3fff
DM_ERROR_EXTRA_INFORMATION (Extra error code information for an above error.)