General Maintenance and Testing
204
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
To test the defibrillator by performing the function checks, you need a simulator. To troubleshoot
device performance, see General Troubleshooting Tips (on page 214).
Standard (Hard) Paddles Check
Perform the standard paddles check as a part of completing the daily Operator’s Checklist that
is provided in the back of this manual.
Function Checks
The following function checks are provided to help personnel keep acquainted with normal
operating procedures and to troubleshoot LIFEPAK 15 monitor/defibrillator performance.
Note:
If your organization downloads device electronic patient records for post-event review,
consider entering "TEST" as the patient’s name to distinguish simulator function tests from
actual patient uses.
Patient ECG Cable Check
Equipment Needed:
•
LIFEPAK 15 monitor/defibrillator
•
Fully charged batteries or power adapter connected to a reliable power source
•
Patient ECG cable (3-lead, 12-lead, or 5-wire)
•
3-lead or 12-lead simulator
To check the patient ECG cable:
1.
Press
ON
.
2.
Connect the ECG cable to the defibrillator.
3.
Connect all cable leads to the simulator.
4.
Turn on the simulator and select a rhythm.
5.
Confirm that Lead
II
is selected.
6.
After a few seconds, confirm that the screen displays a rhythm and that no
LEADS OFF
or
SERVICE
message appears.
7.
For 12-lead cable, press
12-LEAD
and wait for printout. Confirm that a rhythm prints for
each lead.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......