Sterilizable Internal Defibrillation Paddles
158
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Cleaning and Inspecting Pediatric Paddles
Individually protect paddles before and after cleaning to prevent damage to paddle surfaces.
After each use:
1.
Wipe or rinse paddle electrodes, cable connector, paddle handles, and cables with mild
soap and water or disinfectant using a damp sponge, towel, or brush. Do not immerse or
soak.
2.
Dry thoroughly.
3.
Examine paddle surfaces, connector, handles, and cables for damage or signs of wear.
•
Cables that show signs of wear such as loose cable connections, exposed wires, or
cable connector corrosion should be removed from use immediately.
•
Paddles that have rough or pitted electrodes should be removed from use
immediately.
Sterilizable Internal Defibrillation Paddles
Physio-Control internal paddles are specifically designed for open chest cardiac defibrillation.
Figure 44
Sterilizable Internal Defibrillation Paddles
Internal paddles are available in several sizes. To order internal paddles, contact your
Physio-Control representative. In the USA, call Customer Support at 1.800.442.1142, option 2.
For complete information about using internal paddles to provide open chest cardiac
defibrillation, see the
Instructions for Use
provided with the internal paddles.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......