Chapter 5
| Therapy
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
141
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Therapy cable damaged
•
Replace therapy cable and
perform daily checks per
Operator’s Checklist.
REPLACE BATTERY
prompt
occurs
Both batteries are very low
•
Replace one or both
batteries immediately.
•
Connect to auxiliary power
using approved power
adapter.
CPR time shown
(minutes/seconds) is different
than expected
Metronome is on
•
None. The metronome
adjusts the CPR time to
ensure CPR cycle ends with
compressions.
Incorrect setup option selected
•
Change CPR time setup
option. See
LIFEPAK 15 Monitor/Defibrill
ator Setup Options
provided
with your device.
Home Screen is blank but
ON
LED is illuminated
Screen not functioning properly
•
Print ECG strip to assess
rhythm and other active vital
signs.
•
Press
ANALYZE
and use
AED mode, if necessary.
For general troubleshooting tips, see General Troubleshooting Tips (on page 214).
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......