Page | 18
5.1.2
M
EASUREMENT
No device calibration is required before making chlorophyll fluorescence measurement. Results of fluorescence measurement are
affected by the device settings and the physiology of the sample.
Steps for Chlorophyll Fluorescence measurements with AquaPen-C:
•
Fill the cuvette with the sample of algae or cyanobacteria and close the cuvette with the stopper. Fill the cuvette with 3 ml of the
sample. Minimal volume for accurate measurements is 2 ml.
•
Place the cuvette with the sample inside the AquaPen cuvette holder and close the cover to allow dark adaptation.
•
Dark adaptation of the sample is required prior to the following measurements: F
0
, QY, NPQ, LC. Duration of dark-adaptation period
depends on species but mostly varies between 5 and 15 minutes.
•
Mix the sample to avoid sedimentation by holding the AquaPen cover and turning over a few times.
This is essential to prevent
inaccurate readings
.
•
Turn ON the device
–
hold
SET
button for 1 sec.
•
Select
Measurement and press SET
> select required parameter for example - Ft.
•
Press
SET
to start the measurement.
•
AquaPen will measure the parameter. If a protocol was selected such as
OJIP
,
LC
or
NPQ
the display will only show the progress of
the measurements in % but no data will be visible.
•
When measuring Ft and QY the value of the parameter will appear on the display after completion of the measurement. To visualize
the data obtained with OJIP, NPQ or LC protocol recorded data has to be download from the AquaPen to the PC computer via USB
cable or the Bluetooth connection using FluorPen Software (page 44).
•
All measured data are stored in the device memory and can be downloaded to PC after completion of the experiment.
Steps for Chlorophyll Fluorescence measurement with AquaPen-P:
•
For measurements of Ft, QY, NPQ, LC the sample requires dark adaptation period of 5-15 min (this varies with species). Place the
sample in the dark to achieve this. Turn ON the device
–
hold
SET
button for 1 sec.
•
Select
Measurement
and press SET> select required parameter for example -
Ft
.
•
Submerge the probe in the sample and ensure that no air bubbles get trapped inside the probe.
•
Press
SET
to start measurements.
•
AquaPen will measure the parameter. If a protocol was selected such as
OJIP
,
LC
or
NPQ
the display will only show the progress of
the measurements in % but no data will be visible.
•
When measuring Ft and QY the value of the parameter will appear on the display after completion of the measurement. To visualize
the data obtained with OJIP, NPQ or LC protocol, recorded data has to be download from the AquaPen to the PC computer via USB
cable or the Bluetooth connection using FluorPen Software (page 44).
•
All measured data are stored in the device memory and can be downloaded to PC after completion of the experiment.
5.1.3
OJIP
PROTOCOL
The AquaPen device offers the protocol to capture rapid fluorescence transient
–
OJIP, which occurs during exposure of photosynthetic
organisms to high irradiance. The FluorPen software enables data downloading to a PC and subsequent OJIP visualization of the analyzed
data in a graphical and tabular format.
Summary of Contents for AquaPen-C AP 110-C
Page 1: ......
Page 9: ...Page 9 3 DEVICE DESCRIPTION Fig 1 Device description...
Page 13: ...Page 13 Fig 3 AquaPen C AP110 C Fig 4 AquaPen P AP 110 P...
Page 22: ...Page 22 Fig 8 NPQ Protocol...
Page 24: ...Page 24 Fig 9 LC1 Protocol...
Page 25: ...Page 25 Fig 10 LC2 Protocol...
Page 29: ...Page 29...
Page 30: ...Page 30...
Page 31: ...Page 31...
Page 32: ...Page 32...
Page 33: ...Page 33...
Page 34: ...Page 34...
Page 35: ...Page 35...
Page 47: ...Page 47 Fig 26 Data export...
Page 51: ...Page 51 Fig 34 GPS coordinates...