Mounting
The sensor can be mounted to a wall or lie flat on the floor. To mount the sensor to a wall, install
the two drywall anchors (if necessary) and attach the Power Sensor using the two #6 tapping
screws. When installing the sensor on the floor, attach the four rubber feet to the bottom of the
enclosure.
Figure 2: Mounting the sensor
Configuration
All IMS Solution sensors are auto-configured when you plug them into the Host or Node. When a
new sensor is plugged into the Host or Node, the configuration is set to the factory default via the
Sensor Template.
If you have modified these templates, the configuration will be set to the modi-
fied configuration. When a sensor is removed, or the connection is disrupted, the configuration
data remains intact but the sensor will go into trouble status and generate a trouble alarm. When
the sensor is plugged in again, or the connection is re-established, configuration data remains
unchanged and the sensor returns to normal operating condition.
!
Caution
: Removing a sensor for less than 60 seconds does not produce a trouble alarm.
Removing or unplugging a sensor for more than 60 seconds or plugging in a different
sensor at any time will cause a trouble alarm.
Chapter 7: IMS-4000 Sensors
129
Summary of Contents for Sensaphone IMS-4000
Page 1: ...IMS 4000 User s Manual Version 2 4 8 PHONETICS INC SENSAPHONE ...
Page 44: ......
Page 59: ......
Page 60: ......
Page 102: ......
Page 106: ...IMS 4000 Manual 106 ...
Page 135: ...Chapter 7 IMS 4000 Sensors 135 ...
Page 146: ...IMS 4000 Manual 146 ...
Page 148: ...IMS 4000 Manual 148 ...
Page 158: ...IMS 4000 Manual 158 ...
Page 159: ...159 ...