Adjusting Recirculation Options
The following operating parameters can be adjusted:
Initial Rinse
–
The time the system recirculates when restarted before converting to alternating
polishing and standby cycling. Default setting is 120 minutes.
NOTE: Initial rinse can be bypassed to enter Polishing mode by holding the Initial Rinse button at
the bottom of the display for 3 seconds. This is useful if the system is restarted with previously
rinsed consumable packs.
Polish Time
–
The time the system recirculates water to maintain purity. Default setting 2 minutes
Standby Time
–
Dwell time between polishing cycles to conserve resin life. Default setting 10
minutes
Sample Delay
–
The amount of time the system alternates between reading resistivity at the output
of the conditioning pack and the final water quality at the output of the polishing filter. Default
time is 10 seconds.
To adjust settings:
•
Home Screen
–
Press Menu
•
Main Menu
–
Press Configuration
•
Press the appropriate field box; key pad will open
•
Adjust value as required
•
Press ENT to exit key pad
•
Press and hold SAVE until beep is heard
•
Press BACK to return to Configuration Menu
Press to adjust
conditioning
cartridge set point
Press to adjust
polishing filter set
point