System Specifications
Clinical accuracy
The
Gmate™ System is calibrated to yield results
equivalent to plasma glucose concentrations and is
traceable to a NIST standard. The accuracy of the
Gmate™ On System was assessed by comparing
blood glucose results obtained by patients with
those obtained using a YSI 2300 Glucose Analyzer.
System accuracy result for glucose concentrations
< 100 mg/dL(5.5 mmol/L)
System accuracy result for glucose concentrations
≥ 100 mg/dL(5.5 mmol/L)
within ±5 %
90.9% (374/411)
within ±10 %
100% (411/411)
within ±15 %
100% (411/411)
Regression statistics
Slope
0.9857
Intercept
2.9141 mg/dL
R²
0.9976
No. of samples
600
Range tested
34.2-447 mg/dL (1.9 - 24.8 mmol/L)
40
within ±5 mg/dL(0.28 mmol/L)
87.3% (165/189)
within ±10 mg/dL(0.56 mmol/L)
100% (189/189)
within ±15 mg/dL(0.83 mmol/L)
100% (189/189)