Operation
6-17
5. Reconnect the patient circuit to the mask/interface.
6. Select Start Ventilation to initiate ventilation.
Troubleshooting
If Test Failed is displayed, check for leaks in the patient circuit, and install an
exhalation device with lower leak characteristics. Repeat test. If the exhalation
port test fails again, the intentional leak is unknown and Tot.Leak rather than
Pt. Leak is displayed in the patient data window.
Other functions: the
Menu window
From the Menu window you can adjust user preferences.
Brightness
Use Brightness to adjust the screen for optimum daytime or nighttime viewing.
Loudness
WARNING:
WARNING:
Avoid blocking the alarm speakers beneath the ventilator.
Use Loudness to adjust the volume of the alarm and touchscreen audible
feedback. You will hear audible feedback as you go through the selections.
The Alarm Volume Escalation status is also displayed on this screen. See
“Alarm Volume Escalation” on page E-11 for more information.
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......