Operation
6-15
5. Run the exhalation port test if indicated in the table (see “Running
the exhalation port test” on page 6-16 for instructions).
CAUTION:
If you selected PEV or Other as an exhalation port, you must run an
exhalation port test.
NOTE:
If the exhalation port test is not run or if it fails, the intentional leak is
unknown. Tot.Leak rather than Pt. Leak is displayed in the patient
data window.
Table 6-2: Exhalation port selections
Port type
*
Exhalation port test
recommended?
FEP
Philips Respironics Filtered Exhalation Port
No
DEP
Philips Respironics Disposable Exhalation
Port
No
Whisper Swivel
Philips Respironics Whisper Swivel
No
PEV
Philips Respironics Plateau Exhalation Valve
Yes
Other
Exhalation port not supplied by Philips Respi-
ronics.
Yes
None
No inline circuit exhalation port
NOTE:
If you select None, refer to the manufacturer’s
instructions to make sure the mask selected contains an
exhalation port.
No
* Depending on your software version, your available exhalation port selections may vary.
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......