Setting up the ventilator for use
5-11
Using the ventilator
for intra-hospital
transport
WARNING:
Always check the status of the oxygen cylinders before using the
ventilator during transport.
WARNING:
WARNING:
The V60/V60 Plus Ventilator requires a pressurized oxygen supply that
provides a minimum flow of 175 SLPM. Do not use any devices such as
valves, hoses, Grab n' Go regulators or other brands of combined cylinder/
regulators that limit supply of oxygen flow below 175 SLPM.
WARNING:
Do not leave the ventilator unattended when stationed on an incline.
Do the following to conserve oxygen during transport with the ventilator.
•
Make sure all cylinders are full (13,790 kPa/2000 psig or more).
•
Make sure the cylinder regulators are turned off while the ventilator is
connected to wall oxygen.
•
Never turn the cylinder regulator on until you are ready to begin
transport.
•
Only turn one cylinder regulator on at a time. If you turn on both
cylinders, they may become depleted simultaneously, leaving you with
no backup oxygen.
•
Whenever possible, reduce the O
2
setting before transport.
•
Minimize all inadvertent leaks. Tighten masks prior to transport, and
loosen up when patient is back on wall oxygen.
•
Avoid using masks that have an exhalation port built into the mask
when there is already an exhalation port in the circuit.
•
Be aware that oxygen is more rapidly depleted at higher leak rates (see
Figure 5-4).
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......