First-time installation
A-9
Installing the oxygen
manifold kit
If desired, install the oxygen manifold kit as described in the accompanying
instructions.
Verifying ventilator
operation and audible
alarm
Perform the following steps to verify ventilator and audible alarm operation:
1. Assemble and install a patient circuit. (See Chapter 5 for instructions
on installing a patient circuit.)
2. Power on the ventilator and verify that it completes the power-on self-
test.
3. Disconnect the proximal pressure airway pressure line from the
ventilator connector, and verify that the Proximal Pressure Line
Disconnect alarm is annunciated (audio, visual, and flashing alarm
LED).
4. Reconnect the proximal pressure line, and manually reset the alarm.
5. Turn the ventilator off.
6. Remove the patient circuit.
The ventilator is ready to be set up for use as described in Chapter 5.
Configuration and
screen calibration
After completing the setup activities described in Chapter 5, set or check the
ventilator settings for language, units of measure, and time in the diagnostic
mode (see Appendix E). Calibrate the screen as required, referring to
Appendix E
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......