Functional Description of the Monitor Hardware
16
Introduction to the Instrument
Functional Description of the Monitor Hardware
The Monitor receives data from the Measurement Server and Measurement Server Extension
via the Server-to-Monitor link bar and presents this data on the color LCD display. The
following block diagram shows the main functional areas.
AC
Power Supply
AC
Smart Battery
Optional
Vbat,
I2C
LAN
Connector
Bezel Assembly
Keyboard
TouchStrips
IrDA
Standby On/Off (PIC)
4 hardkeys(HIF)
LEDs:
X Bell(HIF)
red alarm(HIF)
yellow alarm(HIF)
On/Off(+5V)
AC Power(PIC)
Battery(PIC)
Display Adapter
Board
Inverter Board
LCD
Display
Backlight
48 Pin
DC/DC Converter
48V current limiter
Battery charger
Display
ECG
Out
SRL to
Measurement
CPU System(360)
(Flash,SRAM,DRAM)
HIF(83C552)
(TouchStrip,Keys,Sound,
LEDs, Alarm Relay,
Battery)
34 Wire
40 Wire
Processor
(Battery
6 Pin
2x2 Pin
31 Pin
Flat Cable
Flat Cable
cable
Flex
cable
Connector
6 Pin
Connector
5 Pin
Connector
48V,
AC present
(Network)
SRL
LAN
Se
ri
al
L
ink
uP bus
Video
I2C
Rx,Tx
T
ouc
hSt
ri
p
Ke
y
s
A
lar
m LEDs
P
IC L
E
D
s,
S
ta
nd
by
Vbat 48V
AC
present
I2C
Board
(Infrared)
Display Assembly
Video
Controller
Controller)
Server
Alarm
Relay VGA
(M3000A)
48Vlim +5V
Vbuf
ECG Out/
Marker In
Controller
System Board
Ventilator
Fan
Wireless LAN
Optional
Connector
140 Pin
or
Serial
Interface
Summary of Contents for M3000A
Page 2: ...Printed in Germany 03 05 Sixth Edition M3046 9300F Part Number M3046 9300F 4512 610 07711 S ...
Page 8: ...vi ...
Page 134: ...Answers to the Tutorial for Installing the Instrument 118 Installing the Instrument ...
Page 268: ...Monitor Parts 252 Replacement Parts Monitor Parts 5 4 3 7 2 1 6 ...